SpO2 Module VM-SpO2

  • Multi Functional Module
  • Cardiac Output, Spo2, 4-Temp, 4-IBP
  • High performance ILIC technology
  • Light Adjust technology
  • Is strong to Moving Artifact
  • Is strong to Low Perfusion

  • Measurement Mode Nellor Mode
    Measurement Range 0 ~ 100%
    Accuracy 70 ~ 100% SpO2 : ± 2%, 50 ~ 69% SpO2 : ± 3%, 0 ~ 49% SpO2 : Unspecified
    Low Perfusion 0.2% to 20%(Clinical Dynamic: SmartSat SpO2 Simulator)
    Setting Time Wave out Time: 2 sec Max, SpO2 Percentage Display: 10 sec Max
    Pulse Range 20 ~ 300bpm
    Pulse Measurement Accuracy 20 ~ 240bpm ±2 bpm, 241 ~ 300bpm ±3 bpm
    Power DC 5V, 350mA
    Dimensions W 135 x D 70 [mm]

  • Channel 4-CH (Option), 2-CH (Standard)
    Measurement Range 10 ~ 45℃
    Accuracy 10.0℃ ~ 24.9℃ ± 0.2℃, 25.0℃ ~ 45.0℃ ± 0.1℃

  • Channel 4-CH (Option), 2-CH (Standard)
    Measurement Range -40 ~ 360 mmHg
    Accuracy ± 1 %
    Sensor Catheters 5uV/mmHg

  • Measurement Mode Thermodilution
    Measurement Range 0.1 ~20L/min
    CO Accuracy 5%
    Injection temperature range 23-43℃
    Catheters Edwards compatible