STANDARDS & OPTIONS
- Multi Functional Module
- Cardiac Output, Spo2, 4-Temp, 4-IBP
- High performance ILIC technology
- Light Adjust technology
- Is strong to Moving Artifact
- Is strong to Low Perfusion
Specifications
SpO2
Measurement Mode | Nellor Mode |
---|---|
Measurement Range | 0 ~ 100% |
Accuracy | 70 ~ 100% SpO2 : ± 2%, 50 ~ 69% SpO2 : ± 3%, 0 ~ 49% SpO2 : Unspecified |
Low Perfusion | 0.2% to 20%(Clinical Dynamic: SmartSat SpO2 Simulator) |
Setting Time | Wave out Time: 2 sec Max, SpO2 Percentage Display: 10 sec Max |
Pulse Range | 20 ~ 300bpm |
Pulse Measurement Accuracy | 20 ~ 240bpm ±2 bpm, 241 ~ 300bpm ±3 bpm |
Power | DC 5V, 350mA |
Dimensions | W 75 x D 70 [mm] |
TEMPERATURE
Channel | 4-CH (Option), 2-CH (Standard) |
---|---|
Measurement Range | 10 ~ 45℃ |
Accuracy | 10.0℃ ~ 24.9℃ ± 0.2℃, 25.0℃ ~ 45.0℃ ± 0.1℃ |
IBP(INVASIVE BLOOD PRESSURE)
Channel | 4-CH (Option), 2-CH (Standard) |
---|---|
Measurement Range | -40 ~ 360 mmHg |
Accuracy | ± 1 % |
Sensor Catheters | 5uV/mmHg |
CARDIAC OUTPUT
Measurement Mode | Thermodilution |
---|---|
Measurement Range | 0.1 ~20L/min |
CO Accuracy | 5% |
Injection temperature range | 23-43℃ |
Catheters | Edwards compatible |